When calibrating metrology equipment, it is
important to remember some specific details about the instrument that you are
working to calibrate. Each dimensional instrument has specifications that must
be met and for some instruments, the calibration must be an ongoing process.
With a Vernier caliper, however, once it has been calibrated, it will
essentially stay calibrated until the end of time... or it is moved out of
alignment somehow.
Here's what you need to remember as you
calibrate your Vernier caliper:
Step #1: Make
sure the jaws of the caliper are clean of any grease, dirt, or foreign debris
that could cause a misalignment. It is critically
important for a Vernier caliper to be clean as it is calibrated. Simple soap
and warm water are often enough to clean one, but sometimes a heavier cleaning
agent must be used, especially for tough grease. Avoid using steel wool.
Step #2: Make
sure that the gear is able to move along the tracks freely. If the gear does not move smoothly, then chances are you've missed
some dirt or debris during the cleaning process. Don't force the gear – simply
take another moment to clean the Vernier caliper.
Step #3: Bring
the jaws completely together. When you've done
this, the reading on the caliper should be zero. If it is not, then this is the
place to reset the dial so that it does read zero.
Step #4: Calibrate
the OD jaws. For the first check on the OD jaws,
insert a .5 inch block between the jaws of the caliper and then close the jaws
so that each side touches the gauge block. Record the reading, then repeat with
a 1 inch block and a 4 inch block.
Step #5: Calibrate
the ID jaws. Repeat the steps as listed in Step #4.
If you have another a calibrated caliper, you can instead compare measurements
of the calipers set at 12.7 millimeters and adjust the non-calibrated caliper
to the calibrated one.
Step #6: Check
the height calibration. Using a surface plate,
utilize a three standard sized gauge blocks and bring the jaws together on each
of them. Record the readings.
You Will Also Need To Calculate
Uncertainty
For many people calibrating their metrology
equipment, Step #6 is the last step and they begin to work. Unfortunately this
means they are still using non-calibrated equipment. When calibrating a Vernier
caliper, it is also necessary to calculate the uncertainty of the calipers.
This usually takes 3 specific readings so that you can determine your first set
of uncertainties.
The second set of uncertainty that must be
calculated is based on environmental conditions. Temperature, humidity, and
even the resolution of the caliper itself can all lead to another type of
uncertainty that must be accounted for during the calibration process.
By remembering these steps, you'll be able
to effectively calibrate any Vernier caliper. Use them today to make sure that
your dimensional instrument is always able to measure up!
No comments:
Post a Comment